6.4.4
Calibration of the pressure sensors
Calibrate the permeate pressure
sensor
Follow the instructions below to calibrate the permeate pressure sensor. The
permeate pressure sensor is optional.
Step
Action
1
Tap
Settings on the main screen.
2
Tap
Calibration to open Calibration Settings.
3
Tap
Permeate pressure (1).
1
2
4
Enter the factory settings for the pressure unit connected to the system, in
the gray fields (2) marked
Zero and Span.
For an instruction how to use set-points, see
Adjust set-points, on page 161
5
Tap
Enter.
Result:
The permeate pressure factory calibration settings for the installed unit are
applied to system.
Calibrate pressure sensors
Contact Cytiva for a full-range calibration for the following sensors:
•
Feed pressure sensor
Pf
•
Permeate pressure sensor
Pp
•
Retentate pressure sensor
Pr