ALARM MARK
Over higher limit - ">" = ASCII code 3 E
H
Under lower limit - "<" = ASCII code 3 C
H
Without alarm - ASCI code 2 Ø
H
SYS, MEAN, DIA, PR – Each value consists of 3 bytes in ASCII code.
YEAR, MONTH, DAY, HOUR, TIME – Each value consists of 2 bytes in ASCII code.
YEAR consists of 2 bytes of most significant digits of the year, e.g. 1992 = "92".
NOTE
: The Mean Blood Pressure value is calculated each measurement, but is not displayed on the front
panel for 8800 MS/MSP only.
Output Terminal DSUB-15
(on rear panel; Corometrics Version–Dip switch No. 4 = OFF)
1. Communication Specification
a. 600 bps
b. 8 data bits
c. None parity
d. 1 stop bit
e. Ready/Busy control by CTS signal (Data can be transmitted from Press-Mate when CTS signal level is
HIGH).
2. Output
When a measurement is completed, data is output according the following format.
B
B
A PRESENT TIME (hr, min, sec)* Ø
Ø
11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29
Ø
1
2
3 4 5 6 7 8 9 10
MEASUREMENT TIME
MEAN
PR
SYS
DIA CR
(hr, min, sec)*
Each value consists of 3 bytes in ASCII code.
*Each value consists of 2 bytes in ASCII code.
11-4