Upon activation, the LCD will display the following:
Directional Message
HD MODE DIRECTIONS
USE LIST - TO SELECT
USE SET TO ENTER
Press SET to continue
Screen #1
As this message shows, you must choose a selection by pressing the LIST
↑
↑
and
↓
↓
buttons to move the cursor,
then press SET to enable that selection. Press SET now to continue to the next screen, which will appear as
follows:
SELECT PROGRAM
Prog: A B C D E
Modify Read Program
Exit to Routine Oper.
Screen #2
Interval Program Selections A, B, and C are factory set programs. Those programs are as follows:
Prog.
A
→
→
Time 4.00 Hr.
Start
→
→
1.00
@
5 Int.
1.00
→
→
2.00
@
15 Int.
2.00
→
→
4.00
@
30 Int.
4.00
→
→
@
60 Int.
Prog.
B
→
→
Time 3.00 Hr.
Start
→
→
1.00
@
15 Int.
1.00
→
→
3.00
@
30 Int.
3.00
→
→
@
60 Int.
Prog.
C
→
→
Time 1.00 Hr.
Start
→
→
1.00
@
10 Int.
1.00
→
→
@
30 Int.
Programs D and E have not been formatted so that you can create those programs in any manner you wish.
The "MODIFY PROGRAM SELECTION" of Screen #2 allows the user to view/modify the Internal Programs.
The "EXIT TO ROUTINE OPERATION " selection allows the monitor to return to routine operation without its
Hemodynamic Mode.
Note:
If no selection is made within one minute, the monitor will exit to routine operation. No Interval
Measurement will be taken until the operator initiates the first measurement, by pressing the Start/Stop
button.
Caution:
The monitor limits the continuous measurement to 5 minute maximum only and 12 measurements
for 1 minute interval measurements. The monitor will automatically set the interval at 5 minutes after
reaching the limiting time.
4-6
5 min. measurement intervals for one hour
15 min. measurement intervals from the first hour to the second
30 min. measurement intervals from the second hour to the fourth
60 min. measurement intervals from fourth until manually stopped
15 min. measurement intervals for one hour
30 min. measurement intervals from the first hour to the third
60 min. measurement intervals from the third until manually stopped
10 min. measurement intervals for one hour
30 min. measurement intervals from the first hour until manually stopped