Status Bar
The
Status Bar
, located at the bottom of the window, provides some useful information,
nominally (from left to right):
•
Operation type
:
it displays the type of operation which is running or which is
presently selected in the
Menu Bar
.
•
Loc En / Rem En
:
it displays if the channels are enabled locally (
Loc En
) or
remotely (
Rem En
). If the channels are disabled, no
information is displayed.
•
V0 I0 / V1 I1
:
it displays if the channels are in the V0 I0 status
(
V0 I0
) or in
the V1 I1 status (
V1 I1
).
•
I
:
the system is in INTERLOCK.
•
K
:
the system is in KILL.
•
T / N
:
it indicates if either the TTL (
T
) or the NIM (
N
) standard were
selected via the relevant switch.
•
t
t
:
this red diamond is displayed if at least one channel is ON (i.e.
at least one channel has the PW parameter set ON).
•
CAEN SY2527
:
it blinks to indicate error of communication with a board
plugged into the crate. If the communication error source is
removed, the error status is reset (i.e. the CAEN SY2527 does
not blink any more) by entering and then exiting the
Crate Map
window.
7.1.9.1 Channel Status
The channel status messages which can be displayed in the
Status
parameter
column
of the
Channels Window
are as follows:
•
Cal Err
:
Calibration Error: at least one of the calibrations for at least
one of the channels is not correct.
•
Up
:
the channel is ramping up.
•
E Trip-Dwn
:
an external TRIP occurred: the channel is ramping down.
•
I Trip-Dwn
:
an internal TRIP occurred: the channel is ramping down.
•
OvC Dwn
:
an Over Current condition occurred: the channel is ramping
down due to the intervention of the user (not for a TRIP
condition).
•
E Dis Dwn
:
an external DISABLE occurred: the channel is ramping down.
•
Dwn
:
the channel is ramping down (only in the case that none of the
E Trip-Dwn
,
I Trip-Dwn
,
OvC Dwn
and
E Dis Dwn
conditions
occurred).
•
E Tripped
:
the channel is off due to an external TRIP.
•
I Tripped
:
the channel is off due to an internal TRIP.
•
Ext-Dis
:
the channel is off due to the fact that the board was disabled
via the HV EN input connector on its front panel.