5.3
Remote Power-On
To power-On the system remotely follow this procedure:
1.
Turn on the MAIN switch located on the rear panel of the crate: the MAIN LED
(orange), located on the front panel (refer to Front Panel figure in the Appendix),
lights up.
2.
Turn the Power-On key, located on the front panel (refer to Front Panel figure in
the Appendix), in the left position (ON REMOTE);
3.
Send a proper signal (refer to § 4.2.2, p.22 for details) through the REMOTE IN
input connector on the front panel.
Following these operations, the front panel LEDs listed below will be alight (refer to the
figure below):
+5
(green LED on):
it indicates the presence of +5 V power supply; if it is
off it indicates that there is a fault.
+12
(green LED on):
it indicates the presence of +12 V power supply; if it
is off it indicates that there is a fault.
-12
(green LED on):
it indicates the presence of -12 V power supply; if it is
off it indicates that there is a fault.
+48
(green LED on):
it indicates the presence of +48 V power supply; if it
is off it indicates that there is a fault.
After the initial check of the system, the
Welcome Screen
of the User Software
Interface will appear on the LCD screen (refer to Fig. 7.3, p.44).
Remote IN
Remote OUT
Fig. 5.1 – Remote IN-OUT connectors