Fig. 7.15 – Clear Alarm pop-up Window
7.1.17
The UTILITY Menu: Status
This command allows to define the condition to assert the GEN signal and light up the
relevant LED (see § 4.3.7).
Fig. 7.16 shows the pop-up window containing a list of quantities which allow to define
the condition for asserting the GEN signal according to the following relation:
GEN = ENABLE
∧∧
(OVC
∨∨
OVV
∨∨
UNV
∨∨
ALWAYS) .
To select any of these quantities insert an asterisk in the relevant field by using the
arrows. Then select OK and press ENTER.
ENABLE
corresponds to enable the GEN signal generation and consequently must be
always selected in order to assert the GEN signal.
OVC
,
OVV
, and
UNV
, are signals for system status monitoring (refer to § 4.3.1, 4.3.2,
4.3.3, p.25), If one or more of them are selected and the ENABLE is selected too, the
GEN signal is asserted TRUE at the occurrence of any of the selected conditions.
ALWAYS
allows to assert the GEN signal TRUE anyway. If it is selected and the
ENABLE is selected too, it asserts the GEN signal TRUE anyway, independently from
the status of the other signals mentioned above (OVC, OVV and UNV).