Detailed procedure for the calibration of the oxidation reduction potential sensor (1-point method only)
The 1-point calibration procedure is used for a quick calibration by adjusting the zero of the measurement graph
with a buffer solution with a known oxidation reduction potential.
→
Immerse the clean probe in the redox
solution: The 8619 displays the
measured DDP value of the solution.
→
Enter the potential difference of the redox
solution (indicated on the bottle).
Calibration
ORP calibration
Mx:pH/ORP
1st point
OK
F4
1st point
465.0 mV
475.0 mV
ORP Calibr. result
Offset:-55.60 mV
The 8619 displays the calibration result
1)
1)
• a possible "warning" message indicates either an error in the solution or the ageing of the probe.
• a possible "error" message indicates that the probe must be replaced.
148
Adjustmentandcommissioning
Type 8619
English