PV
: Choose the process input to simulate. The options available depend on the modules fitted.
VALUE
: Enter a process value selected from the "PV" function above to check output behaviour.
8.13.3 Checking that the outputs are working correctly
The icon
T
is displayed in place of the icon
whenever the correct operation test is run on an output.
During the test, this output no longer reacts, depending on the physical parameter measured.
Refer to section 8.9 to access the "Tests" menu.
Tests
Mx:Outputs
M0:Outputs
AO2:
AO1:
ENTERING
ENTERING
DO1:
OFF/ON
DO2:
OFF/ON
To exit the "Tests" menu, press the dynamic key, "ABORT".
AO1
: Check that current output 1 on the selected module is working correctly by entering a current value and
then selecting "OK".
AO2
: Check that current output 2 on the selected module is working correctly by entering a current value and
then selecting "OK".
DO1
: Check that digital output 1 on the module selected is working correctly by selecting the state "ON" or
"OFF" and then "OK".
DO2
: Check that digital output 2 on the module selected is working correctly by selecting the state "ON" or
"OFF" and then "OK".
157
Adjustmentandcommissioning
Type 8619
English