Removable screw terminals, 9-positions, grey
ME
RE
RE
GD
CG
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
pH measurement electrode
Reference electrode
Equipotential electrode
(recommended)
Temperature sensor
Black
(1)
Translucent
(1)
Strap (not delivered)
FE = functional earth
(1)
Colour of the wires in Bürkert connection cables with order codes 561904, 561905 or 561906.
Fig. 34 : Wiring a pH sensor and a Pt100 or Pt1000 temperature sensor in asymmetrical mode to a "pH/ORP" module
7.3.19 Examples of connections to the "pH/ORP" module
ME
RE
RE
GD
CG
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Removable screw terminals,
9-positions, grey
Oxidation reduction potential
measurement electrode
Reference electrode
Temperature
sensor
Strap (not delivered)
FE = functional earth
Black
(1)
Translucent
(1)
Brown
(2)
White
(2)
Black
(2)
Blue
(2)
Not connected
(1)
Colours of the wires in Bürkert connection cables with order codes 561904, 561905 or 561906.
(2)
Colours of the wires of the Pt1000 sensor with order code 427023 and its Bürkert connection cable with order code 427113.
Fig. 35 : Wiring a redox sensor type 8203 and a Pt100 or Pt1000 temperature sensor to a "pH/ORP" module
64
Installationandwiring
Type 8619
English
Summary of Contents for 8619 multiCELL WM AC
Page 6: ...6 English ...
Page 12: ...12 Type 8619 English ...
Page 24: ...24 Type 8619 English ...
Page 40: ...40 Type 8619 English ...
Page 70: ...70 Installation and wiring Type 8619 English ...
Page 180: ...180 Type 8619 English ...
Page 192: ...192 Type 8619 English ...
Page 193: ......
Page 194: ...www burkert com ...