Graph
zone
Description
Name displayed
in the user
defined "Ux"
view (see
section 8.10.8)
Associated code in
the datalogger (see
section 8.10.19)
Status of the
"ON/OFF"
output (see
section 8.10.23)
C
The conductivity of the fluid has
exceeded the value in the table
USP<645> at the corresponding
temperature.
"> Max."
1
ON (output not
inverted)
B
The conductivity of the fluid is between
the percentage set in the function
"USP Alarm" and the value in the table
USP<645> at the corresponding
temperature.
"USP Alarm"
2
ON (output not
inverted)
A
The conductivity of the fluid is below
the percentage set in the function
"USP Alarm" at the corresponding
temperature.
"OK"
0
OFF (output not
inverted)
CALIB. TEMP.
: Choose the value of the temperature used when calibrating the probe/sensor:
• choose "Auto": the temperature of the fluid is measured by the sensor
• choose "Manual": enter the value of the temperature (in °C) in the next field, e.g. when no temperature sensor is
connected to the module.
CALIBRATION SOLUTION
: Choose the calibration solution used for automatic calibration of the conductivity
sensor.
131
Adjustmentandcommissioning
Type 8619
English