![Bürkert 8619 multiCELL WM AC Operating Instructions Manual Download Page 66](http://html.mh-extra.com/html/burkert/8619-multicell-wm-ac/8619-multicell-wm-ac_operating-instructions-manual_2831935066.webp)
wire colour
signal
Removable screw terminals,
9-positions, grey
translucent
White
ME
RE
RE
GD
CG
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Green
Grey
Green/yellow
Blue
Red
strap (not
delivered)
Black
(coax)
FE = functional earth
Translucent
pH sensor
Red (coax cable
shielding)
reference electrode
Blue
rhodium electrode
Green/Yellow
cable shielding
Grey
sensor body
Green
Pt1000
White
Pt1000
Fig. 38 : Wiring a Bürkert sensor type 8201 or 8203 with integrated Pt100 / Pt1000 temperature sensor, in symmetrical
mode using a Variopin connection cable with order code 554856 or 554857
7.3.20 Wiring the conductivity module "COND"
Removable screw terminals, 9-positions, green
Conductivity
sensor
C+
P+
P-
C-
GD
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Temperature
sensor
FE = functional earth
Fig. 39 : Wiring a resistive conductivity sensor with 2 electrodes and a Pt100 or Pt1000 temperature sensor to a conduc-
tivity module "COND"
66
Installationandwiring
Type 8619
English
Summary of Contents for 8619 multiCELL WM AC
Page 6: ...6 English ...
Page 12: ...12 Type 8619 English ...
Page 24: ...24 Type 8619 English ...
Page 40: ...40 Type 8619 English ...
Page 70: ...70 Installation and wiring Type 8619 English ...
Page 180: ...180 Type 8619 English ...
Page 192: ...192 Type 8619 English ...
Page 193: ......
Page 194: ...www burkert com ...