© 2004,1996 BRUNO INDEPENDENT LIVING AIDS, INC.®
SRE-1540 UL 06-18-2004
32
INSTALLATION
CIRCUIT BOARD DIAGNOSTICS :
RESIDENTIAL
Chirp
.25 Seconds
Short Beep
.5 Seconds
Long Beep
1.5 Seconds
Pause
1 Second
CHANGING THE PCB DIAGNOSTIC MODE
* Turn the Circuit Breaker on the carriage to `OFF'.
* Remove Left Carriage Cover.
* Unit is shipped in Multi-User Diagnostic Mode. Changes are made via the
Number 1 and Number 2 positions on the 4-Ganged DIP Switch.
4-GANGED DIP SWITCH (S1)
Also located on the 4-ganged DIP Switch are Switch No. 3 (LH/RH installation) and Switch No.
4 (coast delay). Refer to the Conversion to Right-Hand Installation instructions for changing
Switch No. 3. The coast delay option (Switch No. 4) has been provided in cases of interference
which may cause intermittent operation. The normal setting is 600 mSec. of coast. Should the
unit lose the remote call/send signal, this can be increased to 900 mSec. by moving Switch No.
4 to the `ON' position.
DIAGNOSTIC MODE
DIP SW NO. 1 DIP SW NO. 2
Multi-User
OFF
OFF
Single-User
ON
OFF
Quiet
OFF
ON
Battery Warning Only
ON
ON
OTHER CIRCUIT BOARD FEATURES
AUDIO REFERENCE
NOTICE:
Before touching anything inside the carriage assembly,
ground yourself by touching an unpainted metal
surface on the unit such as an exposed bolt,
or one of the mounting screws on the electrical panel.
While you work, periodically touch an unpainted metal
surface to dissipate any static electricity that could
harm internal components.