IFS 125M User Manual
Bruker Optik GmbH
Maintenance 6
1
Turn the main power switch in the OFF posi-
tion to switch off the spectrometer.
2
Remove the MCT detector from the spectrometer. Detailed information are
given in chapter 5.10.
3
Connect the connecting piece of the vacuum
adapter to the vacuum pump.
The connecting piece has an outer diameter
of 9.7mm.
4
Pull the knob to the open position.
5
Loosen the coupling nut.
6
Carefully slip the vacuum adapter over the connecting piece of the detector
dewar and fasten the coupling nut finger-tight.
The O-ring must engage properly into the corresponding groove at the
detector port. Additional tightening is not necessary
7
Push the knob into the closed position until
the union nut of the vacuum adapter is in con-
tact with the dewar evacuation valve.
8
• Screw the union nut into the dewar evacua-
tion valve by turning the knob clockwise; 2 to
3 rotations are sufficient. Do not tighten the
union nut completely as this would prevent it
from being unscrewed later.
• Evacuate the vacuum adapter by using the
vacuum pump.
Table 6.16: Evacuating detector
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...