Firmware Update E
During firmware update
• During firmware update, a log window is displayed showing all actions performed
by the FCONF program.
• The log-file is stored in the same directory as the backup files.
• If an error occurs during the update procedure, the FCONF program terminates
the procedure and recommends to restore the previous firmware version.
After firmware update has finished
• If the firmware update has finished, the FCONF program resets the spectrome-
ter. The log window reads:
Resetting the spectrometer... done
.
• After a successful spectrometer initialization, the firmware version is displayed in
the log window.
3
• Activate
Update firmware
option but-
ton.
• Click the
Next
button.
4
• Either accept the default directory
setting or define a different path by
clicking the
Browse
button.
• Click the
Next
button.
5
Click the
Finish
button to start the
update.
i
The update procedure may take sev-
eral minutes, depending on the
available bandwidth and the amount
of files to be updated.
Table B.1:
Updating firmware
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...