184
IFS 125M User Manual
Bruker Optik GmbH
Connection Ports 10
G
CAN bus
These CAN bus connectors are used for external automa-
tion units like microscopes, user-specific switching mirrors
etc. They are identical to the top ones (A) with regard to
their function.
H
LSB122 port
The LSB122 port provides the connection to the laser and
light switch signals of the scanner. It provides a direct
connection to the interferometer compartment.
I
ICON port
The ICON port controls the source.
J
FLAPS port
The FLAPS port is not used with IFS 125M.
K
DDC port 0/1/2/
3
The ports DDC 0 to 3 are versatile ports to connect exter-
nal optical modules and detectors. These ports include a
complete CAN bus, transmit all remote trigger signals
required and establish a complete connection to DDC
compatible detectors. The
DDC
abbreviation means
Digi-
tal Detector Connection
.
Component
Definition
Table 10.2: Ports on flange panel
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...