130
IFS 125M User Manual
Bruker Optik GmbH
Troubleshooting 7
7.5
No interferogram observable in Check Signal mode
In this case, it is assumed that no interferogram will be displayed on the
Check Signal
tab on the OPUS
Measurement
dialog.
Assuming that the spectrometer can be accessed and that there is an optical connection
between the interferometer outlet and detector inlet, the problem can be caused by:
Possible causes
Solutions
No optics connected
See chapter 7.7.
OPUS status light is gray
Check the optical bench URL.
☞
On the OPUS
Measure
menu, select the
Optic Setup and Service
command.
☞
Click the
Optic Bench
tab.
Wrong measurement para-
meters
Use the measurement parameters as described in
appendix C.
Optical path blocked
Check whether any accessory blocks the IR
beam.
Detector not or incorrectly
installed. The following mes-
sage pops up:
Device not
connected.
No analog board
selected
.
In case of manually changed detectors:
• check whether the detector is properly inserted
into its holder.
• check whether the DDC cable is connected
properly (see chapter 3.6.1).
No or wrong detector selected
• On the OPUS
Measure
menu, click the
Advanced Measurement
command
.
• Click the
Optics
tab to select the correct detec-
tor.
Wrong peak position saved
• On the OPUS
Measure
menu, select the
Advanced Measurement
command.
• Click the
Check Signal
tab to determine and
save the correct peak position:
☞
Use the < and > scan range buttons to shift
the start position of the scanner.
☞
After each click, wait until a new interfero-
gram is displayed. Once the peak is found,
click the
Save Peak Position
button.
☞
If the true peak position is too far away from
the saved peak position, determine the true
position from a
Sample Single Channel
mea-
surement as explained in chapter 8.3.3
Table 7.4: No interferogram observable
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...