Operation 5
i
The position value can change any time you power up the spectrometer as the scanner
is always re-initialized during start-up. Thus, after powering up the spectrometer or re-
initializing the scanner, select the
Advanced Measurement
command of OPUS and click
the
Check Signal
tab. Then, click the
Save Peak Position
button.
To verify the currently detected signal intensity, compare the amplitude value displayed
on the
Check Signal
tab with the amplitude value stated in the supplied test protocol
1
.
If there is not any signal detected or if the amplitude value displayed on the
Check Signal
tab deviates significantly from amplitude value given on the supplied test protocol, check
the installation of the spectrometer component(s) you have replaced before.
Definition
A
Amplitude value:
Indicates the currently detected signal intensity. The amplitude is proportional
to the intensity of the incident IR radiation on the detector. It should not have a
short-term deviation of more than 200 ADC counts if the beam is directed onto
the thermally stable calibration source.
Long-term deviation (over several hours) can lead to larger deviations in the
signal intensity due to variations in the ambient temperature. The signal inten-
sity should not deviate by more than 20% from the value recorded during the
delivery in the installation report.
Position:
The position represents the position (in HeNe fringe zero crossings) of the
interferogram maximum. This value must be constant when the OPUS
Check
Signal
mode is selected (as long as the scanner is not stopped or re-initial-
ized).
If it varies, check whether the spectrometer is subject to extreme vibrations. If
there are two interferogram maxima of nearly the same intensity, the position
may vary between two values.
B
Interferogram
It indicates that a signal is detected.
Table 5.4: OPUS check signal tab
1. The supplied test protocol documents the result of a factory-performed test. The test has been performed
with the spectrometer being optimally adjusted. You will find the test protocol in the blue folder supplied with
the spectrometer.
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...