Installation 3
3.2
Packaging
3.3
Transportation
Due to its weight (approx. 120 kg) the spectrometer should be carried by at least 3 per-
sons. When transporting the spectrometer, use the original packaging to avoid dam-
ages.
3.4
Site requirements
The operating company has to provide an installation site that meets the following site
requirements:
CAUTION
Damaged packaging
Injury and/or property damage
➣
Contact shipping company.
➣
Do not put spectrometer into operation.
Environment:
• Temperature range: 18 - 35°C
• Humidity (non-condensing): less than 80% (relative
humidity)
Electromagnetic inter-
ferences:
• The spectrometer is shielded against electromag-
netic interferences.
• Do not locate the spectrometer near potential induc-
tive electrical interference, high energy pulses and
high electric or magnetic fields.
• Use lines which are separated from those which are
connected to machines (e.g. elevators, large electric
motors, etc.) with large electrical surges or bursts.
Table 3.2: Site requirements
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...