72
IFS 125M User Manual
Bruker Optik GmbH
Maintenance 6
5
• Scroll to
Scanner Control
.
• Select
Front Short Adjust Mode
. Write down the peak-to-peak amplitudes
(LASA and LASB) detected on the oscillograph.
• Select
Fast Adjust Mode
to check the signal amplitude over the full path
length.
• Select
Back Short Adjust Mode
.
• Wait until the scanner commutes around the end position.
• Compare the amplitudes detected on the end position with the amplitudes
on the front position. If the amplitude decrease is more than 35%, you have
to readjust the laser (see chapter 6.2.4).
a. Firmware version June 2012 or later display the laser signal amplitudes (see chapter 6.2.2, step 1) in all
adjust modes, too. Therefore, connecting an oscillograph is not mandatory.
b. Alternatively, you can access the spectrometer by the WEB interface. Details are described in
chapter 8.3.5.
Table 6.2: Signal decrease over optical path
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...