124
IFS 125M User Manual
Bruker Optik GmbH
Troubleshooting 7
7.1.1
Remote fault diagnostics via full report
Remote fault diagnostics means that you send a complete spectrometer status report
(Full Report) to Bruker via e-mail. This report enables a Bruker service technician to per-
form a first remote fault diagnostics.
The procedure for sending the report is different and depends on whether your spec-
trometer is connected to a network, network computer or a stand-alone computer.
7.1.1.1
Sending full report if your spectrometer is connected to a network/network
computer
With OPUS version 6 or higher, it is possible to send the full report to Bruker via e-mail.
Proceed as follows:
1.
Click the OPUS status light (chapter 7.2). The
Instrument Status
dialog opens.
2.
Click the
Send Report
button. Clicking this button sends the report to the
opusre-
e-mail address.
i
To be able to use this feature an e-mail program must be installed on the network com-
puter, and an e-mail account set up.
7.1.1.2
Sending full report if your spectrometer is connected to a stand-alone computer
1.
Generate a full report manually and save it (chapter 8.3.4.1).
2.
Transfer the full report file to a network computer. Make sure that an e-mail program
is installed on this network computer, and an e-mail account set up.
3.
Send the full report to
as an attached file.
i
To be able to use this feature an e-mail program must be installed on the network com-
puter, and an e-mail account set up.
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...