IFS 125M User Manual
Bruker Optik GmbH
Data Communication 9
9.3
Assigning network addresses
When delivered, the spectrometer uses the standard IP address 10.10.0.1. Assigning a
different IP address only becomes necessary if the spectrometer is to be directly con-
nected to a network. This IP address has to be assigned by the network administrator.
The FCONF (Firmware Configuration) program is used to assign a new IP address to
the spectrometer. This program is part of the OPUS software. The single steps required
for assigning IP addresses are described in the following:
1
Start FCONF program.
☞
Double click the
fconf.exe
file in the OPUS base path.
2
• Activate the
Assign a new address to
the
spectrometer button.
• Click the
Next
button.
Table 9.4:
Assigning network address
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...