143
Bruker Optik GmbH
IFS 125M User Manual
Web Interface 8
8.3
Spectrometer diagnostics pages
Due to the integrated EWS16 board (Embedded Webserver Board), the spectrometer
operates like a Web server, i.e. the spectrometer can be accessed using a Web browser
(e.g. Microsoft Internet Explorer, Firefox Browser). This kind of spectrometer accessing
requires a physical connection between the spectrometer and PC as well as knowledge
about the spectrometer IP address.
If you open the Web browser and enter the spectrometer IP address (default IP address:
10.10.0.1) into the address field of the Web browser (A in figure 8.2), the spectrometer
home page of the particular spectrometer is displayed. If you do not know the IP
address, see the white label on the mobile electronics unit.
The spectrometer diagnostics pages are a useful and powerful fault diagnostics tool,
especially the
Direct Command Entry
,
Diagnostics
and
Service
pages. Some of these
pages (and their sub-pages, if available) are described in the following chapters.
Figure 8.2:
IFS 125M spectrometer home page
A
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...