Default Parameter Settings
Data acquisition is performed and automated by the OPUS software. The following table
includes parameters recommended for the standard spectrometer configuration (MIR).
i
Depending on the spectrometer configuration, different default parameter settings may
apply.
1.
On the OPUS
Measure
menu, select the
Setup Measurement Parameters
com-
mand.
2.
Enter the following measurement parameters on the different tabs:
Advanced parameters
Settings
Resolution
2
Sample/Background Scans
32
Save Data
0 to 7000
Result Spectrum
Transmittance
Data Blocks to be saved
Single Channel/Sample Interferogram
Optics parameters
Settings
Source Setting
MIR
Beamsplitter
KBr (MIR)
Optical Filter Setting
Open
Aperture Setting
1.1 mm
Detector Setting
LN-MCT [Internal Pos. 1]
Scanner Velocity
40 kHz
Sample/Background Signal Gain
x1
Delay after Device Change
0 sec
Delay before Measurement
0 sec
Table C.1: Default parameter settings
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...