115
Bruker Optik GmbH
IFS 125M User Manual
Maintenance 6
6.8.7
Checking optical alignment
It is important that the laser has been aligned carefully before proceeding with the follow-
ing steps.
1
Put a paper in front of the focussing mirror (as screen for the laser light
which is back-reflected through the aperture).
2
• Select the smallest aperture:
☞
On the OPUS
Measure
menu select
Direct Command
Entry.
☞
Enter
apt=300
.
☞
Click
Send Command
.
• Dim the lights to optimally see the back-reflected laser spot on the paper.
i
The aim of the alignment is to have the back-reflected laser spot passing
through the center of the smallest aperture. This ensures that the aper-
ture is exactly on the optical axis of the interferometer.
Table 6.29: Transporting spectrometer - Checking optical alignment
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...