166
IFS 125M User Manual
Bruker Optik GmbH
Web Interface 8
8.3.4.1
Generating full report
i
Save the full report immediately after a problem or failure has occurred. Otherwise,
important information will be overwritten by newer entries.
8.3.4.2
Interpreting a full report
The full report summarizes the information of the spectrometer diagnostics pages
(chapter 8.3.2 et seq.). It contains all information about the spectrometer. In case of a
spectrometer problem, the full report helps the service technician to quickly narrow down
the cause of the problem.
To quickly locate a source of trouble, simply scroll down the full report until you encoun-
ter a red ERROR field, see figure 8.18. For a more detailed fault diagnostics read
through the individual information provided in the full report.
1
• Open Web browser.
• Enter spectrometer IP address into
the address entry field.
• On configuration page, click
Service
.
2
Click the
Full Report
option. The cur-
rent full report is displayed.
3
• On the
File
menu of the Web browser click the
Save as
command.
• Use the file extension
*.htm
.
• Send original htm file to Bruker. Do not generate any screenshots or text files.
Table 8.14:
Generating full report
Figure 8.18: ERROR field
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...