106
IFS 125M User Manual
Bruker Optik GmbH
Maintenance 6
6.7.7
Remounting covers
1
• Put your hands below the flange of the input port and the diagonally
located hanger.
• Put on the cover of the detector and interferometer block.
➣
Make sure that the two eyelets at the inner edge of the cover are
catched by the two guiding posts which are located on the console.
• Close the clips on each side
a
.
2
• Insert the trough into of the sample compart-
ment.
• Use a flat-tip screwdriver to fasten the two
knurled thumb screws, located on the bot-
tom of the trough.
3
• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid clock-
wise.
➣
Press down the screw and rotate it once.
4
• Insert the Teflon-made counter nuts (A).
i
The Teflon-made counter nuts are only
available if you use a liquid nitrogen-
cooled detector.
• Put on the protective caps (B).
Table 6.25: Transporting spectrometer - Remounting covers
A
B
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...