80
IFS 125M User Manual
Bruker Optik GmbH
Maintenance 6
6.3.2
Handling instructions for tungsten source
• Do not touch the tungsten bulb of the NIR source with your fingers as this con-
taminates the bulb and significantly reduces the service life.
• Wear gloves when changing the bulb and observe the operating instructions
indicated on the source packing.
CAUTION
NIR/VIS source operated out of the spectrometer:
Risk of UV radiation.
➣
Do not operate the NIR/VIS source out of the spectrometer.
➣
Operate according to the safety instructions to avoid risk of UV radiation.