121
Bruker Optik GmbH
IFS 125M User Manual
Maintenance 6
4
• Loosen the clamp screw located between
the two spring-loaded bolts (5 in figure 6.9).
• Adjust screws (4 in figure 6.9), which causes
a x-y shift, so that the observed diffraction
pattern is at maximum brightness.
☞
If required, re-adjust the pre-load by
means of screws (5).
• Adjust screws (1 in figure 6.9), which causes
a tilt, to achieve a (nearly) concentric diffrac-
tion pattern.
☞
If the pre-load of any screw (1) gets too
high or too low, the tension can be cor-
rected by means of the screw (2) which is
diagonally opposite.
• If necessary, the x-y and tilt adjustments
have to be repeated.
• Remove the alignment tool.
5
• Mount the fiber patch cord.
➣
As the fiber is very sensitive, be careful by putting the fiber on the cou-
pler. Do not use too much force and turn the fiber slightly (at the plug)
while guiding the ferrule into the socket.
i
The whole receptacle is mounted at a 3.5° angle, so the ferrule must be
tilted at that angle, too.
• Make sure that the key at the plug points into the direction of the notch at
the receptacle.
• Fix the lock nut only hand-tight.
6
You should see a light beam (at least dim) emitting out of the fiber. You can
watch the beam on a white piece of paper.
• If available, mount the fiber collimator tool with integrated polarizer and Si-
diode at the fiber end.
• Rotate the integrated Si-diode until a maximum intensity is read. Lock the
position. If the tool is not available, use the power-meter or Si-diode to
monitor the intensity (P
out
).
• Turn screws (4) slightly one by one to increase the intensity.
☞
If required, re-adjust the pre-load by screws (5).
• Adjust the tilt of the coupler. Use screws (1) one by one to change the tilt
and maximize the intensity.
☞
If required, re-adjust the pre-load by the screw (2) located diagonally
opposite.
• Repeat the x-y and tilt adjustments successively until you achieve a power
P
out
a
, which is at least 65% of P
in
b
.
Table 6.31: Readjusting fiber-coupling head
Summary of Contents for IFS 125M
Page 1: ...IFS 125M User Manual 1015996 ...
Page 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Page 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Page 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Page 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Page 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Page 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Page 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Page 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Page 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Page 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Page 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...