5 Preparation
38
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Possible foot switch combinations
Foot switch
(type)
Connection
socket
Active output
(front)
Mode
Operation
50
3 or 4
Bipolar Coag
X
Change foot switch output:
Press key/display
44/48
for
1.5 sec. or with foot pedal
sensor (EP+)
2 Monopolar
Coag
X
Change foot switch output:
Press key/display
39
for
1.5 sec. or with foot pedal
sensor (EP+)
Single pedal
with/without sensor
(EP/EP+)
52
3 Bipolar
Coag
X
Change foot switch output:
Press key/display
44
for
1.5 sec. or with foot pedal
sensor (EP+)
50
3 or 4
Bipolar Cut
1)
/Coag
X
Change foot switch output:
Press key/display
44/48
for
1.5 sec. or with foot pedal
sensor (DP+)
Double pedal
with/without sensor
(DP/DP+)
52
2 or 3
Monopolar
Cut/Coag
Bipolar Cut
1)
/Coag
X
Change foot switch output:
Press key/display
39/44
for
1.5 sec. or with foot pedal
sensor (DP+)
1)
Only for ARC 350
The shift sensor on the foot switch is locked in programs 18, 19, and 20
for safety reasons.
Socket changeover must occur directly at the device.
Pos: 34 /679-BOWA/ARC 300/350/6 Vorbereitung/6.5 Funktionsprüfung @ 4\mod_1261040600038_6.doc @ 35014 @ 233
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...