11 Technical specifications
92
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Micro Moderate COAG
7, 8, 10,
11 (up to 30 W)
ARC350MMCOAG_SE_080124_1
0
5
10
15
20
25
30
35
10
100
1000
10000
P [W]
Micro Moderate
COAG 15W
Micro Moderate
COAG 30W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Micro Moderate COAG” =
30 W/15 W
ARC350MMCOAG_SE_080124_2
0
10
5
20
15
25
30
0
10
20
25
15
5
30
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the "Micro Moderate COAG"
setting. Measured load resistance =
125
Ω
ARC350MMCOAG_SE_080125_3
0
20
40
60
80
100
120
140
160
180
200
0
5
10
15
20
25
30
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Micro Moderate
COAG” setting (idle mode)