Guidelines and manufacturer's declaration in
accordance with DIN EN 60601-1-2, Section
6.8.3.201
Electromagnetic interference (DIN EN 60601-1-2, Table 201)
The ARC 300 and ARC 350 are intended for operation in an electromagnetic environment as described below. The
customer or user of the ARC 300 and/or ARC 350 should ensure that these devices are operated in such an
environment.
Interference measurements
Conformity
Electromagnetic environment
guideline
HF emissions as per CISPR 11
Group 2
The ARC 300 and ARC 350 must
emit electromagnetic energy in order
to carry out their intended functions.
Adjacent electronic devices can be
affected.
HF emissions as per CISPR 11
Class A
Emission of harmonics as per
IEC 61000-3-2
Classes A + D
Emission from voltage fluctuations/flicker
as per IEC 61000-3-3
Conforms
The ARC 300 and ARC 350 are
suitable for use in facilities other than
living areas and those connected
directly to the public power supply
network, which also supplies
buildings used for residential
purposes.