7 Detecting and correcting errors
68
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
7.2.
Errors in the EASY monitoring
When working with a split neutral electrode, the following errors may
occur:
EASY monitoring
20
Cause
Indicator
Corrective measures
Flashes green
Significant increase in
resistance
–
Application must not be
terminated.
When the monopolar current
is activated, a significant
problem occurs
An acoustic signal sounds.
A corresponding “INF” - "xxx"
message appears on the 7-
segment indicator
“Monopolar Cut”
2
and the 7-
segment indicator
“Monopolar Coag”
9
.
X
Check the neutral
electrode and the
neutral electrode
cable; see Chapter
EASY neutral
electrode monitoring
(EASY monitoring),
Page 30.
X
Check the neutral
electrode cable for
proper contact and
external damage.
Switches from green to
continuous red
Loosening electrode
An acoustic signal sounds.
The corresponding “INF” –
“xxx” message appears.
The device's electronics
system switches off the
monopolar outputs of the
connection sockets
36
and
40
.
X
Correct the position of
the neutral electrode.
In the case of
continuing error
messages, replace
the neutral electrode.
Pos: 42 /679-BOWA/ARC 300/350/9 Aufbereitung/9 Aufbereitung @ 4\mod_1261040691288_6.doc @ 35166 @ 1
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...