11 Technical specifications
106
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Micro bipolar Neuro COAG
11 (up to 50 W)
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Micro bipolar Neuro COAG”
= 50 W/25 W
•
Diagram of power output P [W] as a
function of the “Micro bipolar Neuro
COAG” setting. Measured load
resistance = 125
Ω
ARC350BNCOAG_SE_080125_2
0
0
10
20
30
40
50
20
40
60
80
100
120
140
160
180
200
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Micro bipolar Neuro
COAG” setting (idle mode)