11 Technical specifications
96
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Micro Spray COAG
7, 10,
11 (up to 30 W)
0
5
10
15
20
25
30
35
10
100
1000
10000
P [W]
ARC350MSCOAG_SE_080125_1
Micro Spray
Coag P=15W
Micro Spray
Coag P=30W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Micro Spray COAG” =
30 W/15 W
ARC350MSCOAG_SE_080125_2
0
5
10
15
20
25
30
0
5
10
15
20
25
30
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the "Micro Spray COAG"
setting. Measured load resistance =
500
Ω
ARC350MSCOAG_SE_080125_3
0
500
1000
1500
2000
2500
3000
3500
0
5
10
15
20
25
30
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Micro Spray COAG”
setting (idle mode)