11 Technical specifications
90
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Moderate COAG Dessication 3
0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
Dessicate 3
60W
Dessicate 3
120W
ARC350MCD3_SE_080124_2
0
20
40
60
80
100
120
140
10
100
1000
10000
R [ ]
P [W]
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Moderate COAG
Dessication 3” = 120 W/60 W
ARC350MCD3_SE_080124_2
0
20
40
60
80
100
120
0
20
40
60
80
100
120
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the “Moderate COAG
Dessication 3” setting. Measured load
resistance = 500
Ω
ARC350MCD3_SE_080125_3
0
100
200
300
400
500
600
0
20
40
60
80
100
120
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Moderate COAG
Dessication 3” setting (idle mode)