The data about settings, points of application, application duration, and
the application of instruments are based on clinical practice. However,
these are only basic guidelines which must be tested and approved for
suitability by the operator. Depending on the individual conditions, it may
be necessary to deviate from the specified data.
Medicine is continuously evolving and growing due to R&D and clinical
practice. These developments may also make deviations from the
specified data necessary.
All programs have basic settings, e.g. operating language, sound or
display and memory options, which can be changed for individual
applications; see Chapter Menu programs, Page 57.
Pos: 37 /679-BOWA/ARC 300/350/7 Bedienung/7.2 Grundeinstellungen der Programme @ 4\mod_1261040627350_6.doc @ 35071 @ 233
6.2.
Basic program settings
6.2.1.
Calling up a program
Call up each of the selectable programs with the following procedure:
1. Press the program key
16/17
repeatedly until the corresponding
program is shown on the display.
2. Connect the instruments; see Chapter Connecting instruments,
Page 35.
6.2.2. Changing
the
program
The preconfigured programs can be selected from "0-23". The
parameters of the preconfigured programs can be changed, and the new
parameters can be saved. Only the special types of current cannot be
changed. To do this, set parameters such as output, blend factor,
coagulation mode and foot switch assignment and then store them by
pressing the AUTOSTART
32
key and keeping it pressed.
The programs from "24-99" can be freely configured. These programs
output standard types of current.
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...