11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
97
Setting
Program
Spray COAG
0-3, 6, 9, 12-20,
4, 5 (only socket 2)
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
Argon Flex Spray COAG I
21-23 (only socket 2)
Argon OPEN Spray COAG II
4, 5 (only socket 1)
ARC350ASCOAG_SE_080125_1
0
20
40
60
80
100
120
140
10
100
1000
10000
P [W]
Argon / Spray
Coag P=60W
Argon / Spray
Coag P=120W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Spray COAG, Argon-Flex
Spray COAG I, Argon-OPEN Spray
COAG II und Argon Ligation” =
120 W/60 W
ARC350ASCOAG_SE_080125_2
0
20
40
60
80
100
120
0
20
40
60
80
100
120
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the setting “Spray COAG,
Argon-Flex Spray COAG I, Argon-
OPEN Spray COAG II”. Measured load
resistance = 500
Ω
ARC350ASCOAG_SE_080125_3
0
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
5500
6000
0
20
40
60
80
100
120
displayed value
U [Vp]
•
Diagram of the HF output voltage
U [Vp] as a function of "Spray COAG“
(idle mode)
ARC350ASCOAG_SE_080125_5
0
500
1000
1500
2000
2500
3000
3500
4000
4500
5000
5500
6000
0
20
40
60
80
100
120
displayed value
U [Vp]
•
“Argon Flex Spray COAG I” program,
connection socket 2
•
Diagram of HF output voltage U [Vp] as
a function of the “Argon Flex Spray
COAG” setting (idle mode)
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...