11 Technical specifications
104
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Bipolar Soft COAG
0, 1, 5, 6, 9, 12-17, 21-99 (socket 3+4),
2, 3, 4, 16, 18-20 (only socket 4)
ARC350BSCOAG_SE_080125_1
0
20
40
60
80
100
120
140
10
100
1000
10000
P [W]
Bipolar Soft
COAG 60W
Bipolar Soft
COAG 120W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Bipolar Soft COAG” =
120 W/60 W
ARC350BSCOAG_SE_080125_3
0
20
40
60
80
100
120
0
20
40
60
80
100
120
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the “Bipolar Soft COAG”
setting. Measured load resistance =
75
Ω
ARC350BSCOAG_SE_080125_2
0
20
40
60
80
100
120
140
160
180
200
0
20
40
60
80
100
120
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Bipolar Soft COAG”
setting (idle mode)