6.3.23.
Program 22: "GastroCUT Pap" (optional for ARC 300)
This program is used for making incisions in papillae with a papillotome
via flexible endoscopes.
In this program, the papillotome is used for cutting and coagulation. Arc
control technology generates the cutting effect with simultaneously
minimized power output.
ARC control facilitates immediate cutting and prevents adhesion of the
electrodes.
The arc is reduced as the blend factor is increased.
Three modes are available in this program: fast, medium, and slow.
For further information on the “GastroCut” mode, see Chapter
"GastroCUT" mode, Page 27.
;
The key/display for “Monopolar output 2”
39
and “Bipolar output 4”
48
are preconfigured.
1. Connect the papillotome to the socket for “Monopolar output 2”
40
(use an adapter if necessary).
2. Connect the foot switch to the HF device.
3. Activate the HF current via the yellow key of the foot switch for
cutting.
–or–
Activate the HF current via the blue key of the foot switch for
coagulation.
4. Set the coagulation level using the keys for scabbing level
5/6
.
5. In the case of hemorrhages:
Press the key/display for “Forced Coagulation” current
13
or use a
clip.
6. If required, remove the papillotome and connect the argon probe
with the argon unit ARC PLUS.
7. Press the key/display for “Spray Coagulation” current”
14
.
8. Press the blue key of the foot switch for coagulation.
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...