6 Operation
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
51
DANGER
Blood loss due to incomplete blood vessel sealing!
X
Deactivate the ligation instrument after completed sealing of blood
vessels.
;
The key/display for “Monopolar output 1”
35
and “Bipolar output 3”
44
are preconfigured.
1. Connect the additional device ARC PLUS.
2. Connect the argon handpiece to “Monopolar output 1”
36
of the HF
device.
–or (only ARC 350L)–
Connect the ligation instrument to “Bipolar output 3”
43
of the HF
device.
6.3.18.
Program 17 "Gynecology Hystero"
This program is used in gynecology for hysterectomies.
6.3.19.
Program 18 "TUR-P Bipolar TUR"
(Bipolar TUR only for ARC 350 and ARC 350L)
•
Both cutting and coagulation are possible in the bipolar
application.
•
Use the connection socket for monopolar instruments
40
for
monopolar TUR.
•
Use the connection socket for bipolar instruments
43
for
bipolar TUR.
•
In the bipolar application, the output is adjusted via the so-
called “effect”. The higher the effect number (0 to 9), the
higher the degree of coagulation.
•
Only coagulation current is available at “Bipolar output 4”
47
.
•
The standard bipolar functions are available for ARC 300 and
ARC 300G.
The shift sensor on the foot switch is locked in this program for safety
reasons.
This program is used for underwater cutting in the transurethral
resection of the prostate.
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...