11 Technical specifications
80
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
11.2.
Output, voltage and current diagrams
Setting
Program
Monopolar Cut
0-5, 16-20,
11 (31 W and above),
23 (only socket 1),
24-99 (socket 1+2)
ARC350MCTUR_SE_080124_1
0
50
100
150
200
250
300
350
10
100
1000
10000
R [
Ω
]
P [W]
Blend 4-9
Blend 3
Blend 2
Blend 0, 1
Measurement of resistances without arcing
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Monopolar Cut” = 300
W/150 W
ARC350MCTUR_SE_080124_2
0
50
100
150
200
250
300
350
10
100
1000
10000
R [ ]
P [W]
Blend 6, 7
Blend 3, 4, 5
Blend 2
Blend 0, 1
Blend 8, 9
Measurement of resistances with prior arcing
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Monopolar Cut” = 300
W/150 W
ARC350MCTUR_SE_080125_3
0
50
100
150
200
250
300
0
0
100
150
200
250
300
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the "Monopolar Cut" setting.
Measured load resistance = 500
Ω
ARC350MCTUR_SE_080125_4
0
100
200
300
400
500
600
700
800
900
1000
0
50
100
150
200
250
300
displayed value
U [Vp]
Blend 6, 7
Blend 3, 4 5
Blend 2
Blend 8, 9
Blend 0, 1
•
Diagram of HF output voltage U [Vp] as
a function of the "Monopolar Cut"
setting (idle mode)
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...