9 Maintenance/Repair
70
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
9. Maintenance/Repair
Pos: 45 /679-BOWA/ARC 300/350/10 Wartung/10.1 Wartung/10.2 Reparatur @ 4\mod_1261040702069_6.doc @ 35223 @ 232
9.1. Maintenance
X
Check the device, the device trolley and the accessories (e.g. foot
switch, cable) after each use for damage or defects. In particular,
make sure that the insulation is intact on all cables.
X
Do not use any damaged device, damaged device trolley or
damaged accessories.
X
Replace defective accessories immediately.
X
Have the safety inspection for the device performed once a year.
9.1.1. Safety
inspection
Safety inspections must be performed once a year.
X
The device and accessories may be inspected only by persons
who have the required training, knowledge or experience and who
can perform the inspection independently.
X
With regard to the safety inspection, you must comply with the
country-specific rules and regulations.
The inspector documents the inspection results and measured values
corresponding to the printed inspection protocol.
In the case of severe deviations from the values of the attached final
acceptance report, or if the specified maximum values were exceeded:
X
Send the HF device to the service center; see Chapter Technical
service, Page 72.
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...