11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
81
Setting
Program
MicroCUT
7, 10,
11 (up to 30 W)
MacroCUT
6, 8, 9
0
5
10
15
20
25
30
35
10
100
1000
10000
P [W]
Blend 9
Blend 8
Blend 7
Blend 6
Blend 5
Blend 4
Blend 3
Blend 2
Blend 1
Blend 0
ARC350MPC_SE_080124_1
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “MicroCUT” = 30 W/15 W
Blend 9
Blend 8
Blend 7
Blend 6
Blend 5
Blend 4
Blend 3
Blend 2
Blend 1
Blend 0
0
10
20
30
40
50
60
70
80
10
100
1000
10000
P [W]
ARC350MPC_SE_080124_2
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “MacroCUT” = 75 W/38 W
ARC350MPC_SE_080124_3
0
10
20
30
40
50
60
70
80
0
10
20
30
40
50
60
70
80
displayed value
P [W]
P
max
Micro Programme
P
max
Macro, Local Anestesia,
HNO
•
Diagram of power output P [W] as a
function of the "MacroCUT" setting.
Measured load resistance = 500
Ω
Blend 8
Blend 7
Blend 6
Blend 9
Blend 4
Blend 3
Blend 2
Blend 5
Blend 0
Blend 1
ARC350MPC_SE_080124_4
0
0
10
20
30
40
50
60
70
100
200
300
400
500
600
700
displayed value
U [Vp]
P MicroCUT
max
•
Diagram of HF output voltage U [Vp] as
a function of the "Blend" setting (idle
mode)
Summary of Contents for ARC 300
Page 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Page 2: ......
Page 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Page 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Page 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...