70
4.13.1.10.4 Sweep STARt
This command sets the start frequency of the sweep:
Arguments
Type:
Numeric.
Units:
MHz, KHz, Hz (default)
Range:
Dependent on the frequency range of the current function.
Rounding:
The value is rounded to 4 digits.
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:STARt<ws><frequency>[units]
[:SOURce:]SWEEP:STARt<ws>MINimum|MAXimum
Examples:
SWEEP:STARt 5KHZ
SWEEP:STARt 5E3
SWEEP:STARt MAXIMUM
SWEEP:STARt MIN
Query
Syntax:
[:SOURce:]SWEEP:STARt?[<ws>MAXimum|MINimum]
Examples:
SWEEP:STARt ?
SWEEP:STARt ? MAX
Response: NR3
4.13.1.10.5 Sweep STOP
This command sets the stop frequency of the sweep:
Arguments
Type:
Numeric.
Units:
MHz, KHz, Hz (default)
Range:
Dependent on the frequency range of the current function.
Rounding:
The value is rounded to 4 digits.
Command Type: Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:STOP<ws><frequency>[units]
[:SOURce:]SWEEP:STOP<ws>MINimum|MAXimum
Examples:
SWEEP:STOP 5KHZ
SWEEP:STOP 5E3
SWEEP:STOP MAXIMUM
SWEEP:STOP MIN
Query
Syntax:
[:SOURce:]SWEEP:STOP?[<ws>MAXimum|MINimum]
Examples:
SWEEP:STOP ?
SWEEP:STOP ? MAX
Response:
NR3
4.13.1.11 PHASe adjust
This command controls the phase adjustment
Arguments
Type:
Numeric
Units:
one (degrees implied)
Range:
-180 to +180 (other values converted to this range)
Rounding:
to integer
Command Type: Setting or Query
Setting