55
-258
Media protected.
An attempt was made to write to protected arbitrary waveform memory.
Device-Specific Errors
An error specific to the device occurred. The DDE bit (bit 3) of the Standard Event Status Register is set.
-315
Configuration memory lost.
Device memory has been lost.
-330
Self-test failed.
-350
Queue overflow.
Error codes have been lost due to more than 10 errors being reported without being read.
Query Errors
A query error indicates that the output queue control has detected a problem. This could occur if either an attempt
was made to read data from the instrument if none was available, or when data were lost. Data could be lost when
a query causes data to be formatted for the controller to be read, and the controller sends more commands without
reading the data.
-410
Query INTERRUPTED.
Data were sent before the entire response of a previous query was read.
-420
Query UNTERMINATED.
An attempt was made to read a response before the complete program message meant to generate
that response was sent.
-430
Query DEADLOCKED.
The input buffer and output queue are full, and the controller is attempting to send more data. In
this case the output queue and input buffers will be cleared. Parsing will resume after the END
message is detected.
-440
Query UNTERMINATED after indefinite response.
A query was received in the same program message after a query requiring an indefinite response
was formatted. Essentially this means that the *IDN? common query and the :ARB:DATA?
query should not be followed by more query messages in the same program message.
System Events
System events have positive valued codes. They are not defined by SCPI, but are specific to the instrument.
Sending the :STATus:PRESet command will disable these events from being reported.
401
Power on
402
Operation complete
The *OPC command has been executed.
Warnings
The execution of some commands might cause an undesirable instrument state. The commands are executed, but a
warning is issued. Sending the :STATus:PRESet command disables reporting of warnings. The existence of these
conditions causes a bit in the Status Questionable Condition register to be set (refer to section 4.13.5.4).
For Model 4076
500
Trigger rate short
510
Output overload