F2: TRIG
-
(Triggered) - Triggers one output cycle of the selected waveform for each trigger event.
F3: GATE
- (Gated) - Triggers output cycles as long as the trigger source asserts the gate signal.
F4:BURST -
(Burst) - Triggers output N cycles for each trigger event, where N ranges from 2 to 999,999.
F5: PHASE
-
Selects the start phase of the signal in non-continuous modes. The range is from -180
˚
to +180
˚
,
with a 0.1
˚ resolution
.
In PHASE mode
F2: SET-ZERO
sets the phase reference to zero when few
instruments are connected to the same external reference and need to be synchronized with different
phase relations.
After selecting the TRIG , GATE or BURST menu, the trigger source menu is available:
Trigger
Menu
F1: MAN
- Selects manual as the trigger source. To trigger the waveform generator,
press this MAN TRIG again.
F2: INT
- Selects the internal trigger generator as the trigger source. Change the
internal trigger rate displayed with the rotary input knob.
F3: EXT
- Selects the external trigger signal as the trigger source. The trigger source
is supplied through the TRIG IN connector.
F4: NBRST -
Selects
the number of burst pulses to be output with each trigger. The N can be changed from 2 to
999,999.
F5:PREV
- Goes back to previous menu
3.6.3 WAVEFORM Key
Displays the waveforms available:
Waveform