30
F4: LEAD-TRAIL
- Selects different Rise and Fall times of the Pulse.
F5:PREV
- Returns to previous menu.
3.6.6 UTILITY Key
Utility
Menu
F1: GPIB
- Selects the GPIB remote mode of operation. After selection the GPIB address can be set to any
value from 1 to 31 using the rotary knob. The value is kept in a nonvolatile memory and used at
power-on. The factory default address is 9. Setting the address to 31 puts the device in the off-
bus state (it will not respond to messages on the GPIB bus).
F2:RS232
- Selects the RS232 remote control mode. After selection, the baud rate can be selected as 2400,
9600, 19200, and 38400. The RS-232 uses 8 bit data, 1 stop bit and no parity.
F3:INTEN
- (Intensity) Selects the intensity of the LCD display. Select a value using the rotary input knob.
Valid numeric values are from 1 to 31. The value is kept in the nonvolatile memory, after a 20
seconds time-out. The lower the value, the dimmer the display.
F3: POWER
- (Power-on default) Selects the power-on default setting. Select a value using the numeric
keypad or the rotary input knob. The selection is effective after a 10 s time-out period. Select
zero (0) to have the waveform generator power on with the factory default settings. Select 50 to
have the waveform generator power-on with the settings it had at the last power-off. Select any
other value in the range from 1 to 49 to have the waveform generator power-on with the settings
that you have saved with
STORE(see SETUP key section 3.6.9)
in the range of 1 to 49.
F5: SUM ON/OFF
- (Summing On/Off) Selects the external summing mode. An analog signal connected to the
Sum In connector in the rear panel of the unit is added to the output signal. A 5Vp-p input signal will
generate a full amplitude output signal.
3.6.7 SWEEP Key
Selects the Sweep Mode and allows the entering of sweep parameters as Sweep Start, Sweep Stop and Sweep
Rate.
To select the sweep mode, press SWEEP, then press the function key that corresponds to the desired Sweep menu
option, as shown: