75
Syntax:
:TRIGger:SOURce?
Response:
MAN|BUS|INT|EXT
4.13.3.3 Burst Count
:TRIGger:BURSt <burst count>
Used to set the number of cycles to be output in the BURST mode. It is not a standard SCPI command.
Arguments
Type:
Numeric
Range:
1 to 999999
Rounding:
to integer value
Command Type: Setting or Query
Setting
Syntax
:TRIGger:BURSt<ws><value>
Examples
:TRIG:BURS 100
:TRIG:BURS MAXIMUM
Query
Syntax:
:TRIGger:BURSt?[<ws>MAXimum|MINimum]
Response:
NR1
Examples:
:TRIG:BURST?
:TRIG:BURS? MAX
4.13.3.4 Internal Trigger Rate
:TRIGger:TIMer <trigger rate>
Sets the rate of the internal trigger.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
1E-6S to 100S
Rounding:
to 4 digits
Command Type: Setting or Query
Setting
Syntax:
:TRIGger:TIMer<ws><value>[units]
:TRIGger:TIMer<ws>MINimum|MAXimum
Examples:
:TRIG:TIM 10E-6
:TRIG:TIM MIN
Query
Syntax:
:TRIGger:TIMer?[<ws>MINimum|MAXimum]
Response:
NR3
Examples:
:TRIG:TIM?
:TRIG:TIM? MIN