page 64
Chapter 5:
Maintenance and Troubleshooting
Sally Sue and Peggy Sue User Guide
Yearly
We recommend Sally Sue and Peggy Sue have annual preventive maintenance performed by an authorized
ProteinSimple representative. Please contact Customer Support to schedule a visit.
Leveling the Instrument
This option allows you to see if the instrument is properly leveled. In the Compass for Simple Western main
menu, select
Instrument
and click
Leveling
. A screen will display showing the level status:
NOTE: If the reading is outside the -0.05 to 0.05 range, you’ll need to level the instrument. Follow the
instructions below or contact ProteinSimple Technical Support or your Customer Service Engineer if you
need help.
1.
Loosen the jam nut (11/16”) on both front feet.
Jam Nut
Foot Nut
Foot