Electrical Requirements
Sally Sue and Peggy Sue need a dedicated, grounded circuit capable of delivering the appropriate current
and voltage for your country. The power requirements for select countries are listed below:
Table 2-3:
Power requirements.
In addition to these requirements, we recommend the grounded circuits terminate at the receptacles, and
receptacles must be located within 10 ft (3 m) of the instrument.
Environmental Requirements
Sally Sue and Peggy Sue like a consistent temperature in the lab. They work best when conditions stay
within these ranges:
Table 2-4:
Environmental requirements.
Region
Volts (AC)
Frequency (Hz)
Amps
US and Canada
120
60
5.5
Europe
240
50
2.7
Japan
100
50/60
6.5
Requirement
Specification
Operating temperature range
18 - 24 °C (64 - 75 °F)
Operating humidity range
20-60% relative, non-condensing